GUI reference
R&S
®
ZNB/ZNBT
445
User Manual 1173.9163.02 ─ 62
Sweep All Channels/Sweeps
If you select "Sweep All Channels",
restarts the sweep in all channels
of the active recall set.
For the DEFAULT
, the number of sweeps in single sweep mode is
channel-specific and can be defined in the table below the separator line. The
"Sweeps" column reflects the
settings of the individual channels. The
"Sweeps" setting above the separator line is unused and disabled.
For remote languages other than DEFAULT, the "Sweeps" setting above the separator
line defines the common number of sweeps that are executed on each channel in sin-
gle sweep mode. The table below the separator line is unused and disabled.
Tip:
In remote control, it is possible to retrieve the results acquired in any of the
sweeps within a single sweep group.
Remote command:
INITiate[:IMMediate]:ALL
INITiate<Ch>[:IMMediate]:SCOPe ALL
[SENSe<Ch>:]SWEep:COUNt
Sweep Active Channel/Sweeps (table)
If you select "Sweep Active Channel", then in single sweep mode
only
restarts the sweep in the active channel.
The "Sweeps" column reflects the
settings of the individual channels.
Remote command:
INITiate<Ch>[:IMMediate]:SCOPe
SINGle
5.11
Cal softtool
The "Cal" softtool provides all functions related to system error calibration, scalar
power calibration, and "SMARTerCal".
Access:
Channel – [Cal]
Cal softtool