GUI reference
R&S
®
ZNB/ZNBT
478
User Manual 1173.9163.02 ─ 62
Remote command:
SOURce<Ch>:POWer<PhyPt>:CORRection[:ACQuire]:VERification:
RESult?
Port Overview
Shows all source ports together with the possible power calibrations.
For each analyzer port P1 ... PN, a "Ref. Receiver", a "Meas. Receiver" and a "Source
Flatness" calibration can be performed.
●
If a "SMARTerCal" is active, additional receiver power calibrations are not allowed
and the R&S
ZNB/ZNBT only offers the possibility to perform "Source Flatness"
calibrations of the related test ports:
Chapter 4.5.7.3, "Combining SMARTerCal with scalar power calibration"
on page 200 for details.
●
If external generators are configured, they appear as additional source ports G1 ...
Gm in the port overview. A source power (flatness) calibration is available at each
generator port.
Select a "Ref. Receiver", "Meas. Receiver", or "Source Flatness" symbol to start the
related calibration. See
"Power Cal dialog – Ref. Receiver"
"Power Cal dialog – Source Flatness"
A green checkmark indicates that the calibration data has been acquired success-
fully. A half circle overlay indicates that only the respective port frequencies are cov-
ered (see
"Calibrate Only Port Frequency"
Remote command:
SOURce<Ch>:POWer:CORRection[:ACQuire]
[SENSe<Ch>:]CORRection:POWer<PhyPt>:ACQuire
SOURce<Ch>:POWer:CORRection:DATA:PARameter<Wv>?
SOURce<Ch>:POWer<PhyPt>:CORRection:DATA:PARameter<Wv>:COUNt?
Apply
Is enabled when a new set of power calibration data has been acquired. The button
applies all available power calibrations to the active channel, aborts the verification
sweeps, and closes the port overview section.
The power calibration state is indicated in the trace list, see
on page 197. Use the functions in the
on page 531 to activate, deactivate, or store power calibrations.
Remote command:
SOURce<Ch>:POWer:CORRection[:ACQuire]
[SENSe<Ch>:]CORRection:POWer<PhyPt>:ACQuire
SOURce<Ch>:POWer:CORRection:DATA
Cal softtool