GUI reference
R&S
®
ZNB/ZNBT
491
User Manual 1173.9163.02 ─ 62
Remote command:
[SENSe:]CORRection:COLLect:AUTO:PORTs:CONNection?
Apply
Is enabled when sufficient data have been acquired for the calibrated ports and stand-
ards and for the power meter. The button starts the calculation of the calibration data
and closes the calibration wizard.
The results are applied to the active channel (or to all channels in the active recall set,
if all channels are calibrated).The current instrument settings are stored with the cor-
rection data.
To avoid incompatibilities, older calibration data is deleted unless it has been transfer-
red into the calibration "Pool" using the
.
Use the
to perform an additional source flatness calibration (see
Chapter 4.5.7.3, "Combining SMARTerCal with scalar power calibration"
Note: Checks during the calculation of correction data
Incompatibilities between the selected calibration type, the standards and the channel
settings can cause the calibration to be inaccurate. The analyzer auto-detects potential
sources of errors and displays appropriate, self-explanatory notice boxes.
Remote command:
[SENSe<Ch>:]CORRection:COLLect:AUTO
[SENSe<Ch>:]CORRection:COLLect:AUTO:TYPE
[SENSe<Ch>:]CORRection:COLLect:AUTO:PORTs
[SENSe<Ch>:]CORRection:COLLect:AUTO:PORTs:TYPE
[SENSe<Ch>:]CORRection:COLLect:AUTO:SAVE
[SENSe<Ch>:]CORRection:COLLect:AUTO:POWer
5.11.1.9
SMARTerCal wizard (Cal Unit, MultiCal)
setup and simultaneous acquisition of multiple SMARTerCals.
Access:
Channel – [Cal] > "Start Cal" > "SMARTerCal" – "Start... (Cal Unit)"
The guided SMARTerCal is analogous to a regular automatic calibration with an addi-
tional power calibration sweep at a single analyzer port. It consists of the following
steps:
1.
"Ports":
Define the SMARTerCals to be performed (test ports, calibration unit,
characterization, calibration type, power meter, power port)
Cal softtool