GUI reference
R&S
®
ZNB/ZNBT
496
User Manual 1173.9163.02 ─ 62
Figure 5-34: Multiple port assignments
Similar functionality as for the "SingleCal" version of the wizard (see
on page 487): select the adequate calibration in the "Cal Type"/"Ports" table to
display and edit the corresponding port assignments.
Start
.
If the configured port assignments are invalid, this action is disabled.
Step 3: SMARTerCal
In the final step, the R&S
ZNB/ZNBT displays a "SMARTerCal" screen that guides the
user through the calibration data acquisition.
The logic is identical to that of the single calibration case (see
5.11.1.10
SMARTerCal wizard (manual)
The "SmarterCal (Manual)" wizard guides you through the setup and execution of a
manual SMARTerCal.
Access:
Channel – [Cal] > "Start Cal" > "SMARTerCal" – "Start... (Manual)"
The wizard proceeds through the following steps:
1.
"Ports and Type":
Select the ports to be calibrated, the calibration type, a connec-
ted power meter and the "Power Port".
2.
"Connectors and Cal Kits":
Specify the connector type and gender of the
selected ports. If necessary, load or change a calibration kit.
3.
"SMARTerCal":
Cal softtool