GUI reference
R&S
®
ZNB/ZNBT
504
User Manual 1173.9163.02 ─ 62
The purpose of the typical result traces "Trc1"" and "Trc2" is to avoid connection errors
and to track hardware problems: if the correct standard type is measured, and every-
thing is properly connected, then the measured traces are expected to be similar to the
typical trace.
The S-Parameter traces are labeled P[
j
_
i
]_<standard type> S
ij
, where
j
indicates the
input (test) port and
i
indicates the output port, e.g. P[1_2]_Unknown_Through S21.
If switch matrices are involved, the label also indicates:
●
which VNA port
b
is switched to the input port
j
and
●
which VNA port
a
is switched to the output port
i
Instead of a single S
ij
trace we have multiple traces S
ij
v
a
t
i
v
b
t
j
. E.g., "S21v1t1v2t2"
means that S21 is measured with VNA port 1 connected to the input port and VNA port
2 connected to the output port.
Power Sweep Diagrams
The "Pmtr<n>(P<m>)" traces are displayed during the "Power" calibration sweep.
Ports, Standards and Power
Allows you to perform the required calibration standard measurements, plus the addi-
tional "Power" measurement.
A green checkmark indicates that the calibration data of a standard or power meter has
been acquired successfully. A green checkmark after the port symbol indicates that the
minimum number of calibration measurements for the port has been performed.
Tip: Optional calibration measurements
●
If the selected calibration kit comprises a Sliding Match, then for every required
Match measurement either the Match or at least three positions of the Sliding
Match must be measured. See
Chapter 4.5.2.3, "Sliding Match standards"
●
For a PTRL calibration, at least one line standard must be measured between any
pair of calibrated ports. See
Chapter 4.5.1.9, "TRL calibration"
Note:
The power calibration sweep is performed at the "Reference Receiver Cal
Power" setting from the scalar power calibration; see
"Reference Receiver Cal Power"
Remote command:
[SENSe<Ch>:]CORRection:COLLect[:ACQuire]:SELected
SOURce<Ch>:POWer:CORRection[:ACQuire]
See also:
[SENSe<Ch>:]CORRection:COLLect:LOAD:SELected
Restart Sweep on Std. Meas.
If this function is active, a new standard measurement initiates a new sweep, starting
at the beginning ("Start") of the sweep range: The sweep points for the calibration
sweep are in ascending order, like for an ordinary measurement.
If "Restart Sweep on Std. Meas." is inactive, the new standard measurement is started
at the current sweep point; the current sweep is continued as a calibration sweep.
Note:
This function has a secondary effect in IDLE mode (i.e. while the calibration sweep is
NOT running):
●
If active, only the selected trace is refreshed.
Cal softtool