GUI reference
R&S
®
ZNB/ZNBT
510
User Manual 1173.9163.02 ─ 62
●
"PASS": all measured reflection parameters are within the specified
, compared to the selected reference
●
"FAIL" otherwise.
[In this case, the validation stops at the first failed reflection measurement.]
Remote command:
Advanced Mode
Switches to
.
Advanced mode
In advanced mode you can configure the validation settings (
) and run or rerun particular measurements (
,
Use "Detect Port Assignment" to detect the current port assignment (after reconnecting
the calibration unit).
Measurement
Selects the measurement to be performed in advanced mode and runs it
Characterization
Selects the calibration unit characterization to be used for the validation.
By default the factory characterization is used.
Remote command:
CALCulate:CALValidate:CHARacteriza
Calibration Standard
Selects the calibration standard to be measured in advanced mode.
Validation Limits
The "Limits..." button opens the "Cal Validation Limits" dialog that allows you to specify
the maximum tolerable deviations from the corresponding S-parameters of the
selected reference
Cal softtool