GUI reference
R&S
®
ZNB/ZNBT
546
User Manual 1173.9163.02 ─ 62
5.11.5.4
METAS Reconnection dialog
While a METAS calibration is active, you can notify the METAS VNA Tools about addi-
tional reconnections and/or cable movements. The latter is only applicable if cables
were configured during the setup of the METAS calibration.
Access:
main toolbar icon or
Channel – [Cal] > "METAS Cal" > "METAS Reconnection"
Both reconnections and cable movements contribute to the overall measurement
uncertainty. In the METAS VNA Tools project created during the
these events are registered in the so-called measurement journal. In addition, the
results measured immediately after the event report are persisted in the project.
The effect of a reconnection depends on the selected
, the effect of a cable
movement depends on the selected
(on their respective models, to be precise).
For the METAS calibration, the VNA firmware reports the minimum possible number of
reconnections (depending on the selected calibration type) and no cable movements.
All other events of these types must be reported manually, using the "METAS Recon-
nection" dialog.
Add New Connection/Add Cable Movement
While a METAS calibration is active, these buttons notify the METAS VNA Tools about
additional connector reconnections and/or cable movements. The latter is only applica-
ble if cables were configured during the setup of the METAS calibration.
Remote command:
[SENSe<Ch>:]CORRection:METas:CABLe:REConnection
[SENSe<Ch>:]CORRection:METas:CABLe:MOVement
Measurement Filename
Allows you to configure the name of the measurement data file generated with the
measurement journal entry. For uniqueness, a timestamp is always appended.
Cal softtool