GUI reference
R&S
®
ZNB/ZNBT
618
User Manual 1173.9163.02 ─ 62
If "Adjust Time Gate" is on, the time gate is moved to left (right) when the offset-correc-
ted reference plane is moved to the right (left). Its absolute position remains fixed. With
this setting, it is possible, e.g., to keep the time gate at the position of the antenna con-
nector while the antenna is measured at different length offsets.
Remote command:
CALCulate:FILTer[:GATE]:TIME:AOFFset
Auto Length
Calculates the electrical length offset for the receive port of the active trace. See
If "Delay" is the selected trace format, the entire trace is shifted in vertical direction and
centered on zero. In phase format, the "Auto Length" corrected trace shows the devia-
tion from linear phase.
Use
to limit the frequency range for the calculation.
Note
that in order to calculate the electrical length offset of port
i
, you have to select a
trace that uses port
i
as one of its receive ports. It is
not
relevant for the "Auto Length"
If the measured quantity is a ratio, or if it is derived from a ratio, its receiving port is
given as the index of the wave quantity in the numerator. If the active trace shows an
S-parameter S
ij
, then "Auto Length" adds a length offset at port i. If the receive port is
balanced, then the same offset is applied to both its physical ports.
Remote command:
[SENSe<Ch>:]CORRection:EDELay:AUTO
Fixture Compensation...
The "Fixture Compensation..." button opens the
.
Offset Embed softtool