GUI reference
R&S
®
ZNB/ZNBT
647
User Manual 1173.9163.02 ─ 62
Background information
Refer to
Chapter 4.6.2.7, "Ground loop de-/embedding"
If the "Fixture Simulator" is disabled for the related channel (see
on page 549), this tab is inactive, i.e. all controls except the "Overview" button are
grayed out.
Overview
See
Type
Switches between "Deembedding" and "Embedding" network definition.
All Deembedding Activated
Deactivates or activates all deembeddings (single-ended, balanced, port set, ground
loop) for the active channel, if any.
This control is only visible if
Network
The graphical list contains all available 1-port networks:
The symbol selects "no network" and disables ground loop de-/embedding.
The "1-Port Data" network is defined by imported S-parameter data; see
The remaining networks (Shunt L and Shunt C) are defined by lumped elements
whose parameters are displayed below the graphical list.
Tip:
●
Drag and drop the network symbols in horizontal or vertical direction to switch to
the next symbol.
●
Conductance in Embedding Networks
switch to change from resistances
in "Capacitor in parallel with resistor" model to conductances and vice versa.
Offset Embed softtool