GUI reference
R&S
®
ZNB/ZNBT
656
User Manual 1173.9163.02 ─ 62
Remote command:
CALCulate:FMODel:DEASsistant:DUT:TYPE
CALCulate:FMODel:DEASsistant:PRESet
Left: Model A / Right: Model A|B
Allows you to select a coupon type, i.e. to define how the coupling at the left side (lead-
in) and right side (lead-out) of the DUT is modeled.
If you select the same model for the left and right side and check
on page 657, then also the right side is labeled with "Model A" and in
the measurement phase only the data of a single coupon ("Coupon A") must be mea-
sured or loaded from file.
"Single Ended" This coupon model is appropriate for one or more single-ended inputs
(outputs), where:
●
All inputs (outputs) use the same coupon model and
●
There is no interaction between inputs (outputs) if more than one
is present on the DUT
Figure 5-66: n single-ended input ports (uncoupled)
"Balanced"
This coupon type is appropriate for one or more balanced inputs (out-
puts), if:
●
All inputs (outputs) use the same coupon model and
●
There is no interaction between inputs (outputs) if more than one
is present on the DUT
Figure 5-67: n balanced input ports
Offset Embed softtool