GUI reference
R&S
®
ZNB/ZNBT
689
User Manual 1173.9163.02 ─ 62
Figure 5-74: TDR Stimulus – Advanced Settings dialog
For additional settings, see
Chapter 5.5.5, "Time Domain tab"
Source/Receiver Settings
The "RF Power" is the output power of the R&S
ZNB/ZNBT, "IF Bandwidth" and "Aver-
age Factor" determine the operation of the related R&S
ZNB/ZNBT receiver. For a
description and related remote control commands, see
DC Value
Defines how to calculate the DC gain of the DUT (since the R&S
ZNB/ZNBT cannot
measure down to 0 Hz). These parameters are the same settings as for the standard
time domain option R&S
ZNB/ZNBT-K2. For a description and related remote control
5.16.2.4
Eye Diag tab
The "Eye Diag" tab is for enabling and configuring the eye diagram and its measure-
ments.
Access:
System – [Applic] > "TDR" > "Eye Diag"
Applic softtool