Command reference
R&S
®
ZNB/ZNBT
918
User Manual 1173.9163.02 ─ 62
CALCulate:FMODel:SFD:COUPon:MEASure:FILename
<String>
For a fixture modeling with the SFD tool, this command loads the coupon properties
from a Touchstone file.
The coupon type can be selected using
CALCulate:FMODel:SFD:COUPon:TYPE
.
Parameters:
<String>
Path to the Touchstone file, either absolute or relative to the
selected working directory (
Manual operation:
See
"Load File / 1x Open Preset / 1x Short Preset"
CALCulate:FMODel:SFD:COUPon:MEASure:FILename:CLEar
Invalidates the file path previously set using
.
Usage:
Event
Manual operation:
See
"Load File / 1x Open Preset / 1x Short Preset"
CALCulate:FMODel:SFD<Ph_pt>:COUPon[:STATe]
<Boolean>
For a fixture modeling with the SFD tool, this command allows to specify the ports to
which the test coupon is connected.
This has to be defined before measuring the test coupon (using
).
Suffix:
<Ph_pt>
.
Physical port number
Parameters:
<Boolean>
1 (ON, true) if the test coupon is connected to port <Ph_pt>, 0
(OFF, false) otherwise
Manual operation:
See
CALCulate:FMODel:SFD:COUPon:TYPE
<CouponType>
Sets/gets the coupon type to be measured for a fixture modelling with the SFD tool.
Parameters:
<CouponType>
SYMMetric2x | OPEN1x | SHORt1x
Manual operation:
See
CALCulate:FMODel:SFD:DIFFcfg
<SFDPortConfig>
Tells the SFD tool about the port ordering of the 2x Thru test coupon.