Command reference
R&S
®
ZNB/ZNBT
947
User Manual 1173.9163.02 ─ 62
Suffix:
<Chn>
.
Channel number used to identify the active trace
<Mk>
Marker number (ignored for reference markers)
Parameters:
<Active>
Boolean
Manual operation:
See
CALCulate<Chn>:MARKer<Mk>:FORMat
<OutFormat>
CALCulate<Chn>:MARKer<Mk>:REFerence:FORMat
<OutFormat>
Sets/queries the output format for the (complex) value of the related marker.
Suffix:
<Chn>
.
Channel number used to identify the active trace
<Mk>
Marker number (ignored for reference markers)
Parameters:
<OutFormat>
MLINear | MLOGarithmic | PHASe | POLar | GDELay | REAL |
IMAGinary | SWR | LINPhase | LOGPhase | IMPedance |
ADMittance | DEFault | COMPlex | MDB | MLPHase |
MDPHase | MIMPedance | PIMPedance | PADMittance |
MADMittance | MPIMpedance | MPADmittance | INDX
DEFault
means the reference marker is formatted according
the related trace's default marker format (see
CALCulate<Chn>:MARKer:DEFault:FORMat
For the other marker formats, see
on page 107.
*RST:
DEFault
Manual operation:
See
CALCulate:MARKer:FUNCtion:BWIDth:GMCenter
<arg0>
Defines how bandfilter searches calculate the center frequency of the passband or
stopband.
Parameters:
<arg0>
ON
– use
geometric
mean of lower and upper band edge
OFF
– use
arithmetic
mean
*RST:
n/a (*RST does not affect the calculation rule; the
factory setting is ON/geometric mean)
Manual operation:
See
"Geometric Calculation of Bandfilter Center"
SCPI command reference