Command reference
R&S
®
ZNB/ZNBT
951
User Manual 1173.9163.02 ─ 62
Example:
Suppose that the active recall set contains an active trace no. 1.
CALC:MARK ON
Create marker M1 and assign it to trace no. 1.
CALC:MARK:FUNC:EXEC MAX; RES?
Move the created marker to the absolute maximum of the trace
and query the stimulus and response value of the search result.
Usage:
Setting only
Manual operation:
See
The analyzer provides the following search modes:
Mode
Find...
MAXimum
Absolute maximum in the search range (see
)
MINimum
Absolute maximum in the search range
RPEak
Next valid peak to the right of the current marker position
LPEak
Next valid peak to the left
NPEak
Next highest or lowest value among the valid peaks (next peak)
TARGet
Target value (see
CALCulate<Chn>:MARKer<Mk>:TARGet
)
RTARget
Next target value to the right of the current marker position
LTARget
Next target value to the left
BFILter
Regular markers
only
Bandfilter search. The results are queried using
MMAXimum or
MMINimum
Multiple peak search
SPRogress
Sweep progress
CALCulate<Chn>:MARKer<Mk>:FUNCtion:RESult?
CALCulate<Chn>:MARKer<Mk>:REFerence:FUNCtion:RESult?
Returns the result (stimulus and response value) of a search started by means of
CALCulate<Chn>:MARKer<Mk>:FUNCtion:EXECute
or
MARKer<Mk>:REFerence:FUNCtion:EXECute
. The search must be executed
before the command is enabled.
Suffix:
<Chn>
.
Channel number used to identify the active trace
<Mk>
Marker number (ignored)
Example:
See
CALCulate<Chn>:MARKer<Mk>:FUNCtion:EXECute
Usage:
Query only
Manual operation:
See