Command reference
R&S
®
ZNB/ZNBT
1000
User Manual 1173.9163.02 ─ 62
Options:
R&S
ZNB/ZNBT-K3
Manual operation:
See
CALCulate:TRANsform:TIME...
The
CALCulate:TRANsform:TIME...
commands control the transformation into the
Chapter 4.7.2, "Time domain analysis"
CALCulate<Chn>:TRANsform:TIME:CENTer
.................................................................. 1000
CALCulate<Chn>:TRANsform:TIME:DCHebyshev
.......................................................... 1001
CALCulate<Chn>:TRANsform:TIME:LPASs
....................................................................1001
CALCulate<Chn>:TRANsform:TIME:LPASs:DCSParam
...................................................1001
CALCulate<Chn>:TRANsform:TIME:LPASs:DCSParam:CONTinuous
............................... 1002
CALCulate<Chn>:TRANsform:TIME:LPASs:DCSParam:EXTRapolate
.............................. 1003
CALCulate<Chn>:TRANsform:TIME:LPFRequency
......................................................... 1003
CALCulate<Chn>:TRANsform:TIME:RESolution:EFACtor
................................................ 1003
CALCulate<Chn>:TRANsform:TIME:SPAN
.....................................................................1003
CALCulate<Chn>:TRANsform:TIME:STARt
.................................................................... 1004
CALCulate<Chn>:TRANsform:TIME:STATe
.................................................................... 1004
CALCulate<Chn>:TRANsform:TIME:STIMulus
................................................................1005
CALCulate<Chn>:TRANsform:TIME:STOP
.....................................................................1005
CALCulate<Chn>:TRANsform:TIME[:TYPE]
................................................................... 1006
CALCulate<Chn>:TRANsform:TIME:WINDow
.................................................................1006
CALCulate<Chn>:TRANsform:TIME:XAXis
.....................................................................1006
CALCulate<Chn>:TRANsform:TIME:CENTer
<CenterTime>
Defines the center time of the diagram in time domain.
Suffix:
<Chn>
.
Channel number used to identify the active trace
Parameters:
<CenterTime>
Center time of the diagram in time domain.
Range:
-99.99999999999 s to +99.99999999999 s
Increment: 0.1 ns
*RST:
1.5E-009 s
Default unit: s
Example:
*RST; :CALC:TRAN:TIME:STAT ON
Reset the instrument, activating a frequency sweep, and enable
the time domain transformation for the default trace.
CALC:TRAN:TIME:CENT 0; SPAN 5ns
Set the center time to 0 ns and the time span to 5 ns.
Manual operation:
See
"Time Start / Time Stop / Time Center / Time Span"
Note:
If the x-axis is scaled in distance units (
CALCulate<Chn>:TRANsform:TIME:
DISTance
), then the center value is entered in m; the range and default value
changes accordingly.
SCPI command reference