Parameters:
<Span>
Time span of the diagram in time domain.
Range:
2E-012 s to 200 s.
Increment: 0.1 ns
*RST:
5E-009 s
Default unit: s
Example:
See
CALCulate<Chn>:TRANsform:TIME:CENTer
Manual operation:
See
"Time Start / Time Stop / Time Center / Time Span"
Note:
If the x-axis is scaled in distance units (
CALCulate<Chn>:TRANsform:TIME:
DISTance
), then the span is entered in m; the range and default value
changes accordingly.
CALCulate<Chn>:TRANsform:TIME:STARt
<StartTime>
Defines the start time of the diagram in time domain.
Suffix:
<Chn>
.
Channel number used to identify the active trace
Parameters:
<StartTime>
Start time of the diagram.
Range:
-100 s to +99.999999999998 s.
Increment: 0.1 ns
*RST:
-1E-009 s
Default unit: s
Example:
*RST; :CALC:TRAN:TIME:STAT ON
Reset the instrument, activating a frequency sweep, and enable
the time domain transformation for the default trace.
CALC:TRAN:TIME:STAR 0; STOP 10 ns
Set the start time to 0 ns and the stop time to 10 ns.
Manual operation:
See
"Time Start / Time Stop / Time Center / Time Span"
Note:
If the x-axis is scaled in distance units (
CALCulate<Chn>:TRANsform:TIME:
DISTance
), then the start value is entered in m; the range and default value
changes accordingly.
CALCulate<Chn>:TRANsform:TIME:STATe
<Boolean>
Determines whether the time domain transformation for trace no.
<Chn>
is enabled.
Suffix:
<Chn>
.
Channel number used to identify the active trace
SCPI command reference