Command reference
R&S
®
ZNB/ZNBT
1047
User Manual 1173.9163.02 ─ 62
CALCulate<Ch>:TRANsform:VNETworks:SENDed:DEEMbedding<PhyPt>:
PARameters:C<Cmp>
<CircuitModel>, <Capacitance>
CALCulate<Ch>:TRANsform:VNETworks:SENDed:DEEMbedding<PhyPt>:
PARameters:C<Cmp>?
<CircuitModel>
Specifies the capacitance value C<Cmp> in the different circuit models for single
ended port embedding.
In the query form, the <Capacitance> parameter must be omitted. The command
returns the capacitance value for the specified circuit model.
Suffix:
<Ch>
.
Channel number
<PhyPt>
Physical port number
<Cmp>
Number of capacitance in circuit model. The total number of
capacitances depends on the selected circuit model.
Parameters:
<Capacitance>
Capacitance C<Cmp> for the specified circuit model.
Range:
-1mF to 1 mF.
Increment: 1 fF (1E-15 F)
*RST:
1 pF (1E-12 F)
Default unit: F
Parameters for setting and query:
<CircuitModel>
CSL | LSC | CSC | SLC | SCL | SCC | SHLC | GSL | LSG |
GSG | SLG | SGL | SGG
Possible circuit models (character data); see
Example:
*RST; :CALC:TRAN:VNET:SEND:DEEM:PAR:C2? CSC
Query the default capacitance C2 for the Serial C, shunt C cir-
cuit model. The response is 1E-012 (1 pF).
CALC:TRAN:VNET:SEND:DEEM:PAR:C2 CSC, 2.2E-12
Increase the capacitance to 2.2 pF.
Manual operation:
See
CALCulate<Ch>:TRANsform:VNETworks:SENDed:DEEMbedding<PhyPt>:
PARameters:DATA
<Interchange>, <arg1>
Defines an embedding network for a single-ended port based on the given S-Parame-
ter traces.
Use
●
CALCulate<Ch>:TRANsform:VNETworks:SENDed:DEEMbedding<PhyPt>:
TNDefinition
to select the adequate circuit model
before
executing this com-
mand.
●
MMEMory:LOAD:VNETworks<Ch>:SENDed:DEEMbedding<PhyPt>
cuit data from a Touchstone file located at the R&S
ZNB/ZNBT's file system.
SCPI command reference