Command reference
R&S
®
ZNB/ZNBT
1093
User Manual 1173.9163.02 ─ 62
Manual operation:
See
CONTrol<Ch>:RFFE:TEST:SENSe:TRIGger
Starts the voltage/current measurements on all RFFE pins.
The measurement time can be defined using
.
Suffix:
<Ch>
.
Channel number
This suffix is ignored and can be omitted.
Usage:
Event
Options:
R&S ZN-B15/Z15 Var. 03
Manual operation:
See
CONTrol<Ch>:RFFE:TEST:TIME
<MeasTime>
Sets the measurement time for the voltage/current measurements on the RFFE and
GPIO pins.
Same functionality as
Suffix:
<Ch>
.
Channel number
This suffix is ignored and can be omitted: the measurement time
is valid for all channels.
Parameters:
<MeasTime>
Measurement (= sampling = averaging) time
Range:
95 μs to 100 ms
*RST:
100 ms
Default unit: s
Options:
R&S ZN-B15/Z15 Var. 03
Manual operation:
See
CONTrol<Ch>:SEGMent<Nr>:DELay
<Delay>
For segmented sweeps this command allows to introduce delays between subsequent
sequence steps and between the final sequence step and the sweep (segment) start.
The command/switch sequences are defined using
CONTrol<Ch>:SEGMent<Nr>:GPIO<Port>:
.
Suffix:
<Ch>
.
Channel number
<Nr>
Segment number
SCPI command reference