Command reference
R&S
®
ZNB/ZNBT
1094
User Manual 1173.9163.02 ─ 62
Parameters:
<Delay>
Delay time
Default unit: s
Manual operation:
See
"Wait (Sweep Sequencer Table)"
CONTrol<Ch>:SEGMent<Nr>:GPIO<Port>:VOLTage
<Voltage>
For segmented sweeps this command allows to define the GPIO voltage settings to be
applied at the start of each segment.
The GPIO ports can be enabled/disabled using
. Complementary RFFE commands can be defined using
SEGMent<Nr>:RFFE<Bus>:COMMand:DATA
.
Suffix:
<Ch>
.
Channel number
<Nr>
Segment number
<Port>
GPIO port number
Parameters:
<Voltage>
Range:
-6 to +12 V
Increment: 0.005 V
Default unit: V
Manual operation:
See
"GPIO columns (sweep sequencer table)"
CONTrol<Ch>:SEGMent<Nr>:RFFE<Bus>:COMMand:DATA
<Command>
For segmented sweeps this command allows to define the RFFE command(s) to be
executed at the start of each segment.
Command execution on an RFFE interface can be enabled/disabled using
CONTrol<Ch>:RFFE<Bus>:SETTings[:STATe]
. Complementary GPIO switches
can be defined using
CONTrol<Ch>:SEGMent<Nr>:GPIO<Port>:VOLTage
Suffix:
<Ch>
.
Channel number
<Nr>
Segment number
<Bus>
RFFE bus interface number
Parameters:
<Command>
3 to 37
hexadecimal digits
(0-F), defining the command to be
executed: digit 1 is the slave address, digits 2 and 3 specify the
command number and the remaining digits represent the data
part (up to 17 digit
pairs
).
Manual operation:
See
"RFFE columns (sweep sequencer table)"
SCPI command reference