Command reference
R&S
®
ZNB/ZNBT
1115
User Manual 1173.9163.02 ─ 62
Manual operation:
See
DISPlay[:WINDow<Wnd>]:TRACe<WndTr>:DELete
Releases the assignment between a trace and a diagram area, as defined by means of
DISPlay[:WINDow<Wnd>]:TRACe<WndTr>:FEED
<TraceName>
and expressed by
the
<WndTr>
suffix. The trace itself is not deleted; this must be done via
CALCulate<Ch>:PARameter:DELete
<TraceName>
.
Suffix:
<Wnd>
.
Number of an existing diagram area (defined by means of
ON
).
<WndTr>
Trace number used to distinguish the traces of the same dia-
gram area
<Wnd>
.
Example:
CALC4:PAR:SDEF 'Ch4Tr1', 'S11'
Create channel 4 and a trace named Ch4Tr1 to measure the
input reflection coefficient S
11
.
DISP:WIND2:STAT ON
Create diagram area no. 2.
DISP:WIND2:TRAC9:FEED 'CH4TR1'
Display the generated trace in diagram area no. 2, assigning the
trace number 9 to it.
DISP:WIND2:TRAC9:DELete
Release the assignment between trace no. 9 and window no. 2.
The trace can still be referenced with its trace name Ch4Tr1.
Usage:
Event
DISPlay[:WINDow<Wnd>]:TRACe<WndTr>:EFEed
<TraceName>
CALCulate<Ch>:PARameter:SDEFine
<TraceName>
)
to a diagram area
<Wnd>
to assign the trace to a diagram area using a numeric suffix
DISPlay[:WINDow<Wnd>]:TRACe<WndTr>:Y:OFFSet
command).
Tip:
You can open the "Trace Manager" dialog to obtain an overview of all channels
and traces, including the traces that are not displayed.
Suffix:
<Wnd>
.
Number of an existing diagram area (defined by means of
ON
).
<WndTr>
Trace number. This suffix is ignored; the trace is referenced by
its name.
Setting parameters:
<TraceName>
String parameter for the trace name, e.g. 'Trc4'.
SCPI command reference