DISPlay[:WINDow<Wnd>]:TRACe<WndTr>:X:OFFSet
<StimulusOffset>
Shifts the trace
<WndTr>
in horizontal direction, leaving the positions of all markers
unchanged.
Suffix:
<Wnd>
.
Number of an existing diagram area (defined by means of
ON
).
<WndTr>
Existing trace number, assigned by means of
WINDow<Wnd>]:TRACe<WndTr>:FEED
Parameters:
<StimulusOffset>
Stimulus offset value.
The range and unit depends on the sweep type.
Default unit: NN
Example:
*RST; :DISP:WIND:TRAC:X:OFFS 1MHZ; :DISP:WIND:
TRAC:Y:OFFS 10
Create the default trace and shift it horizontally by 1 MHz, verti-
cally by 10 dB.
Manual operation:
See
DISPlay[:WINDow<Wnd>]:TRACe<WndTr>:Y:OFFSet
<MagnitudeFactor>[,
<PhaseFactor>[, <RealPart>[, <ImaginaryPart>]]]
Modifies all points of the trace
<WndTr>
by means of an added and/or a multiplied
complex constant. The response values M of the trace are transformed according to:
Suffix:
<Wnd>
.
Number of an existing diagram area (defined by means of
ON
).
<WndTr>
Existing trace number, assigned by means of
WINDow<Wnd>]:TRACe<WndTr>:FEED
Parameters:
<MagnitudeFactor>
Multiplied magnitude factor
Range:
-300 dB to + 300 dB
*RST:
0 dB
Default unit: dB
<PhaseFactor>
Multiplied phase factor, optional for setting command but
returned by query
Range:
-3.4*1038 deg to +3.4*1038 deg
*RST:
0 deg
Default unit: deg
<RealPart>
Real and imaginary part of added complex constant, optional for
setting command but returned by query
Range:
-3.4*1038 to +3.4*1038
*RST:
0