Command reference
R&S
®
ZNB/ZNBT
1123
User Manual 1173.9163.02 ─ 62
<TraceName>
Optional string parameter for the trace name, e.g. 'Trc4'. If this
optional parameter is present, both numeric suffixes are ignored
(trace names must be unique across different channels and win-
dows).
Example:
CALC4:PAR:SDEF 'Ch4Tr1', 'S11'
Create channel 4 and a trace named Ch4Tr1 to measure the
input reflection coefficient S
11
.
DISP:WIND2:STAT ON
Create diagram area no. 2.
DISP:WIND2:TRAC9:FEED 'CH4TR1'
Display the generated trace in diagram area no. 2, assigning the
trace number 9 to it.
DISP:WIND2:TRAC9:Y:RPOS 50
or:
DISP:WIND2:TRAC:Y:RPOS 50, 'CH4TR1'
Set the reference position to the center of the diagram area.
Manual operation:
See
DISPlay[:WINDow<Wnd>]:TRACe<WndTr>:ZOOM:BOTTom
<LowEdge>[,
<TraceName>]
DISPlay[:WINDow<Wnd>]:TRACe<WndTr>:ZOOM:TOP
<UppEdge>[,
<TraceName>]
These commands specify the lower and upper edge of the zoom window. In contrast to
manual control, all or part of the zoom window may be outside the original diagram.
Suffix:
<Wnd>
.
Number of an existing diagram (defined by means of
ON
). This suffix is
ignored if the optional
<TraceName>
parameter is used.
<WndTr>
Existing trace number, assigned by means of
WINDow<Wnd>]:TRACe<WndTr>:FEED
. This suffix is ignored if
the optional
<TraceName>
parameter is used.
Parameters:
<UppEdge>
Lower or upper edge of the zoom window. Range and unit
depend on the measured quantity, see
Default unit: NN
<TraceName>
Optional string parameter for the trace name, e.g. 'Trc4'. If this
optional parameter is present, both numeric suffixes are ignored
(trace names must be unique across different channels and win-
dows).
Example:
See
DISPlay[:WINDow<Wnd>]:TRACe<WndTr>:ZOOM[:
Manual operation:
See
SCPI command reference