Command reference
R&S
®
ZNB/ZNBT
1257
User Manual 1173.9163.02 ─ 62
*RST:
0
Manual operation:
See
"Offset > Calculate after De-/Embed."
[SENSe<Ch>:]CORRection:LOSS<PhyPt>
<LossAtFrequency>
[SENSe<Ch>:]CORRection:LOSS<PhyPt>:SECond
<LossAtFrequency>
These commands define the
f
1
and
f
2
.
The reference frequencies can be set using
[SENSe<Ch>:]CORRection:LOSS<PhyPt>:
.
Suffix:
<Ch>
.
Channel number
<PhyPt>
Physical port number
Parameters:
<LossAtFrequency> Loss at reference frequency
Range:
-200 dB to +200 dB
Increment: 0.001 dB
*RST:
0 dB
Default unit: dB
Manual operation:
See
"Loss at DC / Loss at 1st Freq / 1st Freq for Loss / Use 2nd
Freq / Loss at 2nd Freq / 2nd Freq for Loss"
[SENSe<Ch>:]CORRection:LOSS:AUTO
<Activate>
) function to the active trace of channel <Ch>.
Use
CALCulate<Chn>:STATistics:DOMain:USER
commands to limit the fre-
quency range for the calculation.
Suffix:
<Ch>
.
Channel number
Setting parameters:
<Activate>
ONCE
Required constant
Example:
*RST; :CORR:LOSS:AUTO ONCE
Reset the instrument and apply the "Auto Length and Loss"
function to the default trace (Trc1 in channel 1).
Usage:
Setting only
Manual operation:
See
SCPI command reference