Command reference
R&S
®
ZNB/ZNBT
1258
User Manual 1173.9163.02 ─ 62
[SENSe<Ch>:]CORRection:LOSS<PhyPt>:FREQuency
<RefFreq>
[SENSe<Ch>:]CORRection:LOSS<PhyPt>:SECond:FREQuency
<RefFreq>
These commands define the reference frequencies
f
1
and
f
2
for the
mation.
The corresponding losses can be defined using
.
Suffix:
<Ch>
.
Channel number
<PhyPt>
Physical port number
Parameters:
<RefFreq>
Reference frequency
Range:
Frequency range of the analyzer model.
Increment: 1 MHz
*RST:
1000000000 Hz (= 1 GHz)
Default unit: Hz
Manual operation:
See
"Loss at DC / Loss at 1st Freq / 1st Freq for Loss / Use 2nd
Freq / Loss at 2nd Freq / 2nd Freq for Loss"
[SENSe<Ch>:]CORRection:LOSS<PhyPt>:OFFSet
<OffsetLoss>
Defines the frequency-independent part (DC value) of the offset loss.
Suffix:
<Ch>
.
Channel number of the offset-corrected channel
<PhyPt>
Port number of the analyzer
Parameters:
<OffsetLoss>
Frequency-independent part of the offset loss
Range:
-200 dB to +200 dB
Increment: 0.001 dB
*RST:
0 dB
Default unit: dB
Example:
See
[SENSe<Ch>:]CORRection:EDELay<PhyPt>:ELENgth
Manual operation:
See
"Loss at DC / Loss at 1st Freq / 1st Freq for Loss / Use 2nd
Freq / Loss at 2nd Freq / 2nd Freq for Loss"
[SENSe<Ch>:]CORRection:LOSS<PhyPt>:SECond:STATe
<Boolean>
Defines whether the second reference frequency
f
2
approxi-
mation.
The second reference frequency and its associated loss value can be defined using
[SENSe<Ch>:]CORRection:LOSS<PhyPt>:SECond:FREQuency
and
[SENSe<Ch>:]CORRection:LOSS<PhyPt>:SECond
.
SCPI command reference