Command reference
R&S
®
ZNB/ZNBT
1267
User Manual 1173.9163.02 ─ 62
Suffix:
<Chn>
.
Channel number used to identify the active trace
Usage:
Query only
Manual operation:
See
[SENSe<Chn>:]CORRection:SSTate?
Returns the system error correction state label of the active trace in channel
<Chn>
.
The response is a string variable containing the calibration state label in the trace list
('CAL', 'CAI', 'CAL OFF' ..; see
Chapter 4.5.4, "Calibration labels"
Suffix:
<Chn>
.
Channel number used to identify the active trace
Example:
See
Usage:
Query only
Manual operation:
See
7.3.14.9
[SENSe:]FREQuency...
The
[SENSe:]...
commands set frequency-related parameters, especially the mea-
surement and display ranges for the different sweep types. The frequency ranges for
the different instrument models are listed below; for more details refer to the data
sheet.
Table 7-17: Frequency ranges of R&S
ZNB/ZNBT analyzers
Frequency settings
Start, Stop
Center
Span
R&S
ZNB4
9 kHz to 4.5 GHz
>9 kHz to <4.5 GHz
1 Hz to 4.499991 GHz
R&S
ZNB8 /
R&S
ZNBT8
9 kHz to 8.5 GHz
>9 kHz to <8.5 GHz
1 Hz to 8.499991 GHz
R&S
ZNB20 /
R&S
ZNBT20
100 kHz to 20 GHz
>100 kHz to <20 GHz
1 Hz to 19.9999 GHz
R&S
ZNB26 /
R&S
ZNBT26
100 kHz to 26.5 GHz
>100 kHz to <26.5 GHz
1 Hz to 26.4999 GHz
R&S
ZNB40 variant 72
10 MHz to 40 GHz
>10 MHz to <40 GHz
1 Hz to 39.99 GHz
R&S
ZNB40 variants 82,
83 and 84
R&S
ZNBT40
100 kHz to 40 GHz
>100 kHz to <40 GHz
1 Hz to 39.9999 GHz
R&S
ZNB43
100 kHz to 43.5 GHz
>100 kHz to <43.5 GHz
1 Hz to 43.4999 GHz
SCPI command reference