Command reference
R&S
®
ZNB/ZNBT
1269
User Manual 1173.9163.02 ─ 62
Parameters:
<CenterFreq>
Center frequency of the sweep
Increment: 0.1 kHz
Default unit: Hz
Example:
*RST; :SYST:FREQ? MIN; :SYST:FREQ? MAX
Query the frequency range of the analyzer.
FREQ:CENT 100MHz
Set center frequency to 100 MHz.
FREQ:SPAN 50000
Set frequency span to 50 kHz.
Manual operation:
See
"Start Frequency / Stop Frequency / Center Frequency /
Note:
The measurement range defined by means of the center frequency and the cur-
rent span (
) must not exceed the allowed frequency
range of the analyzer. If necessary, the span is reduced to min (<CenterFreq>
–
f
MIN
,
f
MAX
–
<CenterFreq>).
[SENSe<Ch>:]FREQuency:CONVersion
<ConversionMode>
Enables a frequency conversion measurement mode for channel
<Ch>
.
Suffix:
<Ch>
.
Channel number
Parameters:
<ConversionMode>
FUNDamental | ARBitrary | MIXer | IMODulation
FUNDamental
Measurement of the fundamental signal (non frequency-convert-
ing mode). This selection cancels all port-specific frequency and
power settings (see example).
ARBitrary
Frequency conversion mode (arbitrary port frequencies). This
mode is automatically activated when a port-specific frequency
is defined.
MIXer
Mixer mode, configured using
[SENSe<Ch>:]FREQuency:CONVersion:MIXer...
com-
mands.
IMODulation
Intermodulation measurement mode, configured using the
[SENSe<Ch>:]FREQuency:IMODulation...
commands.
*RST:
FUNDamental
Example:
See
SOURce<Ch>:FREQuency<PhyPt>:CONVersion:
Manual operation:
See
SCPI command reference