Command reference
R&S
®
ZNB/ZNBT
1297
User Manual 1173.9163.02 ─ 62
MANual
Enables the manual configuration of independent AGC settings
for each segment, drive port and receiver. See
*RST:
LDIS
Example:
See
Manual operation:
See
[SENSe<Ch>:]POWer:LOSuppress[:STATe]
<Boolean>
Activates/deactivates "LO Suppression" for channel <Ch>.
"LO Suppression" lowers the LO power of all receivers and thus the power leaking out
of the test ports.
Suffix:
<Ch>
.
Channel number
Parameters:
<Boolean>
OFF (0)
LO suppression active
ON (1)
LO suppression inactive
*RST:
OFF
Manual operation:
See
7.3.14.14
[SENSe:]ROSCillator...
The
[SENSe:]ROSCillator...
commands control the frequency reference signal.
[SENSe:]ROSCillator:EXTernal:FREQuency
................................................................... 1297
[SENSe<Ch>:]ROSCillator[:SOURce]
............................................................................ 1298
[SENSe:]ROSCillator:EXTernal:FREQuency
<ExtClockFreq>
Specifies or queries the frequency of the external reference oscillator.
Parameters:
<ExtClockFreq>
Frequency of the external reference clock signal.
Range:
See the data sheet of your analyzer.
*RST:
10 MHz
Default unit: Hz
SCPI command reference