Command reference
R&S
®
ZNB/ZNBT
1330
User Manual 1173.9163.02 ─ 62
Parameters:
<Voltage>
DC voltage range. The entered numeric value is rounded up to
±
300 mV,
±
3 V,
±
20 V. Values above +20 V are rounded down.
Default unit: V
Example:
SENSe1:DC2:RANGe 1; :DC2:RANGE?
Select a DC measurement at DC INPUT2 with an expected DC
voltage range of
±
1 V. The value is rounded up; the response is
3
.
Manual operation:
See
[SENSe<Ch>:]PORT<PhyPt>:ZREFerence
<RealPart>[, <ImaginaryPart>]
Specifies the complex reference impedance for the physical port numbered
<PhyPt>
(impedance renormalization).
Use
[SENSe<Ch>:]LPORt<LogPt>:ZDEFault[:STATe]
to toggle between config-
ured and default reference impedances.
Suffix:
<Ch>
.
Channel number
<PhyPt>
Physical port number
Parameters:
<RealPart>
Real part of the port impedance.
Range:
1 mΩ to 10 MΩ
*RST:
Default reference impedance of the connector fam-
ily assigned to the port (real impedance, e.g. 50 Ω).
Default unit: Ohm
<ImaginaryPart>
Imaginary part of the port impedance. This part may be omitted
to define a real impedance.
Range:
- 10 MΩ to 10 MΩ
*RST:
Default reference impedance of the connector fam-
ily assigned to the port (real impedance, e.g. 50 Ω).
Default unit: Ohm
Example:
PORT2:ZREF 52, 2
Specify a complex reference impedance of 52 Ω + j * 2 Ω for the
(physical) port no. 2.
CALC:TRAN:IMP:RNOR PWAV
Select renormalization of port impedances according to the
power waves theory.
Manual operation:
See
"Single Ended Mode / Common Mode / Differential Mode"
SCPI command reference