Command reference
R&S
®
ZNB/ZNBT
1348
User Manual 1173.9163.02 ─ 62
SOURce:GROup:SIMultaneous:FOFFset:MOFFSet:
DVALue 0 MHz
Set minimum frequency offset for frequency shifted parallel mea-
surement to 0 MHz.
:SOURce:GROup:SIMultaneous:FOFFset:MOFFset:MODE
BANDwidth
:SOURce:GROup:SIMultaneous:FOFFset:MOFFset:
BWFactor 100
Set the minimum frequency offset for frequency shifted parallel
measurement to at least 100 times the measurement bandwidth
Manual operation:
See
"Parallel Measurement with Frequency Offset"
7.3.15.3
SOURce:POWer:CORRection:COLLect...
The
SOURce:POWer:CORRection:COLLect...
commands allow you to set up a
scalar source power calibration.
SOURce:POWer:CORRection:COLLect:AVERage[:COUNt]
<NoReadings>
Sets a limit for the number of calibration sweeps in the source power calibration. The
command is valid for all channels, ports and external generators.
Equivalent command:
SOURce:POWer:CORRection:NREadings
Parameters:
<NoReadings>
Number of readings
Range:
1 to 100
*RST:
2
Example:
See
SOURce<Ch>:POWer:CORRection[:ACQuire]
equivalent command
Manual operation:
See
"Flatness Cal – Max Iterations"
SOURce:POWer:CORRection:COLLect:AVERage:NTOLerance
<Tolerance>
Specifies the maximum deviation of the measured power from the target power of the
calibration. The command is valid for all channels and calibrated ports.
Parameters:
<Tolerance>
Tolerance value
Range:
0.001 dB to 1000 dB
*RST:
1 dB
Default unit: dB
Example:
See
SOURce<Ch>:POWer:CORRection[:ACQuire]
Manual operation:
See
SCPI command reference