Command reference
R&S
®
ZNB/ZNBT
1350
User Manual 1173.9163.02 ─ 62
PMONly
Power meter only, i.e. only the power meter is used during cali-
bration.
To enable a reference reveiver and/or source flatness calibra-
tion, set
SOURce:POWer:CORRection:COLLect:RRECeiver
and/or
SOURce:POWer:CORRection:COLLect:FLATness
ON
.
RRAFter
First the reference receiver will be calibrated by a single power
meter measurement, then the source flatness calibration will be
performed using the calibrated reference receiver.
Requires
SOURce<Ch>:POWer<PhyPt>:CORRection:COLLect:
FLATness ON
.
RRONly
Reference receiver only. The reference receiver will be used to
perform a source flatness calibration. Ideally the reference
receiver was previously calibrated using a power meter.
Requires
SOURce<Ch>:POWer<PhyPt>:CORRection:COLLect:
RRECeiver OFF
.
*RST:
RRAFter
Example:
See
SOURce<Ch>:POWer:CORRection[:ACQuire]
Manual operation:
See
"Flatness Cal – Pwr Cal Method"
SOURce:POWer:CORRection:COLLect:RRECeiver
<Boolean>
Enables or disables calibration of the reference receiver together with the source
power calibration.
Note:
It is not possible to disable flatness calibration (see
on page 1349) and reference receiver calibration
simultaneously.
Parameters:
<Boolean>
Disables (OFF|1) or enables (ON|1) reference receiver calibra-
tion.
*RST:
ON|1
Example:
See
SOURce:POWer:CORRection:COLLect:FLATness
SOURce:POWer:CORRection:COLLect:VNETworks
<DeEmbNwType>[,
<DeEmbType>]
Defines whether the source flatness cal uses the configured offset de-/embedding to
get the power values at the DUT (instead of the calibrated reference plane).
SCPI command reference