Command reference
R&S
®
ZNB/ZNBT
1364
User Manual 1173.9163.02 ─ 62
SOURce<Ch>:POWer:CORRection:GENerator<Gen>[:STATe]
<Boolean>
Enables or disables the source power calibration for channel
<Ch>
and for an external
generator number
<Gen>
. The command is disabled unless a source power calibration
for the external generator has been performed (
). To enable or disable a source power calibration for an analyzer port use
SOURce<Ch>:POWer<PhyPt>:CORRection:STATe
.
Suffix:
<Ch>
.
Calibrated channel number
<Gen>
Generator number
Parameters:
<Boolean>
Enables (ON) or disables (OFF) the source power calibration for
port number
<PhyPt>
.
*RST:
OFF
Example:
See
SOURce<Ch>:POWer:CORRection[:ACQuire]
Manual operation:
See
SOURce<Ch>:POWer<PhyPt>:CORRection:IMODulation:LTONe[:ACQuire]
Starts the source calibration for the lower tone (1
st
power calibration step for intermo-
dulation measurements), stores and applies the calibration data. The external power
meter used is selected via
SOURce<Ch>:POWer<PhyPt>:CORRection:PMETer:ID
.
Suffix:
<Ch>
.
Calibrated channel number
<PhyPt>
Port number This suffix is ignored; the analyzer calibrates the
port that is selected as a source port for the intermodulation
measurement
(
[SENSe<Ch>:]FREQuency:IMODulation:LTONe
).
SCPI command reference