Command reference
R&S
®
ZNB/ZNBT
1415
User Manual 1173.9163.02 ─ 62
Tip:
Switching off the display speeds up the measurement. This command may have
an impact on the update of trace and channel settings; see
.
Parameters:
<Activate>
ON | OFF - switch the display on or off. If the display is switched
on, the analyzer shows the diagrams and traces like in manual
control.
ONCE - switch the display on and show the current trace. This
parameter can be used for occasional checks of the measure-
ment results or settings. The measurement is continued, how-
ever, the measurement results are not updated. Compared to
the
ON
setting,
ONCE
does not slow down the measurement
speed.
Example:
SYST:DISP:UPD ON
Switch the display on to view the traces and diagrams.
SYSTem:DISPlay:XLABels
<Boolean>
Enables or disables the display of X-axis grid labels in cartesian diagrams with linear
scale.
Parameters:
<Boolean>
ON (1)
Show X-axis grid labels
OFF (0)
Don't show X-axis grid labels
*RST:
OFF
Manual operation:
See
SYSTem:ERRor:ALL?
Queries and at the same time deletes all entries in the error queue.
The entries consist of an error number and a short description of the error. Positive
error numbers are instrument-dependent. Negative error numbers are reserved by the
SCPI standard; see section
Error messages and troubleshooting
Example:
SYST:ERR:ALL?
Query all entries in the error queue.
0,"No error"
is returned
if the error queue is empty.
Usage:
Query only
SYSTem:ERRor:COUNt?
Gets the current number of entries in the SCPI error queue.
Usage:
Query only
SCPI command reference