Command reference
R&S
®
ZNB/ZNBT
1437
User Manual 1173.9163.02 ─ 62
sists of TTL pulses of configurable duration and polarity which can be sent before or
after each sweep or measurement point.
Figure 7-1: Output trigger signal at EXT TRIG OUT
TRIGger:CHANnel<Ch>:AUXiliary:DURation
.................................................................. 1437
TRIGger:CHANnel<Ch>:AUXiliary[:ENABle]
................................................................... 1437
TRIGger:CHANnel<Ch>:AUXiliary:INTerval
.....................................................................1438
TRIGger:CHANnel<Ch>:AUXiliary:OPOLarity
................................................................. 1438
TRIGger:CHANnel<Ch>:AUXiliary:POSition
....................................................................1438
TRIGger:CHANnel<Ch>:AUXiliary:DURation
<TrigOutDuration>
Specifies the width of the output trigger pulses at Trigger Out.
The trigger duration must be shorter than the selected sweep interval (see
CHANnel<Ch>:AUXiliary:INTerval
).
If the trigger duration is too long, you can modify one of the following settings:
●
Reduce the trigger duration, if possible.
●
Slow down the sweep: select a smaller bandwidth or increase the "Sweep Time"
setting.
Suffix:
<Ch>
.
Channel number
Parameters:
<TrigOutDuration>
Pulse width between 1
μs (1E-6
s) and 1
s.
*RST:
1 us
Default unit: s
Example:
See
TRIGger:CHANnel<Ch>:AUXiliary[:ENABle]
TRIGger:CHANnel<Ch>:AUXiliary[:ENABle]
<EnableTriggOut>
Turns the external trigger output EXT TRIG OUT for channel no. <Ch> on or off.
Suffix:
<Ch>
.
Channel number
SCPI command reference