Command reference
R&S
®
ZNB/ZNBT
1448
User Manual 1173.9163.02 ─ 62
Indicates the start of the binary block
<Digits>
Decimal value
Gives the number of decimal digits used for the
<Length>
value
<Length>
Decimal value
Number of bytes the follow in the
<Binary data>
part
<Binary data>
Binary data in ASCII format
Example:
//Return security information
DIAG:HUMS:SEC?
#3133{"antimalware":{"name":
"Windows Defender","enabled":false,"upToDate":
true,"timestamp":
"2018-02-08T10:09:22Z"},"firewallEnabled":true}
Usage:
Query only
DIAGnostic:HUMS:SERVice?
Queries the service information of the instrument. Depending on the set data format,
the queried data is either displayed in
XML
or
JSON
format. For more information about
Return values:
<ServiceInfo>
<block_data>
Service information of the connected instrument as block data in
a comma-separated list:
#blockdata{<lastdate>,<requiredservice>,<calibration>:
{<lastcalibration>,<nextDue>,<nextDueExpired>,
<recommendedCalibrationInterval>},<uptime>}
Binary block data with the following syntax:
#<Digits><Length><Binarydata>
#
Indicates the start of the binary block
<Digits>
Decimal value
Gives the number of decimal digits used for the
<Length>
value
<Length>
Decimal value
Number of bytes the follow in the
<Binary data>
part
<Binary data>
Binary data in ASCII format
HUMS and service date commands