CALCulate<Chn>:LIMit:RDOMain:FORMat
<UnitRef>
Deletes the existing limit line and (re-)defines the physical units of the response values
of the limit line. The units of the stimulus values are defined via
.
Tip:
This command is complemented by
CALCulate<Chn>:LIMit:RDOMain:SPACing
Suffix:
<Chn>
.
Channel number used to identify the active trace.
Setting parameters:
<UnitRef>
COMPlex | MAGNitude | PHASe | REAL | IMAGinary | SWR |
GDELay | L | C
Keyword for the physical unit of the response values. The
parameters form four groups: COMPlex, REAL, IMAGinary, and
SWR select dimensionless numbers (U) for the limit line. MAG-
Nitude selects relative units (dB). PHASe selects phase units
(deg). GDELay selects time units (s). L selects inductance units
(H/Henry). C selects capacitance units (F/Farad).
Usage:
Setting only
CALCulate<Chn>:LIMit:RDOMain:SPACing
<Format>
Deletes the existing limit line and (re-)defines the physical units of the response values
of the limit line. The units of the stimulus values are defined via
.
Tip:
This command is complemented by
CALCulate<Chn>:LIMit:RDOMain:FORMat
Suffix:
<Chn>
.
Channel number used to identify the active trace.
Setting parameters:
<Format>
LINear | LOGarithmic | DB | SIC
Keyword for the physical unit of the response values.
Default unit: dB (irrespective of the parameter selected)
Usage:
Setting only
CALCulate<Chn>:LIMit:LOWer:STATe
<Boolean>
CALCulate<Chn>:LIMit:UPPer:STATe
<Boolean>
These commands switch the lower and upper limit check on or off. Lower limit line seg-
ments are assigned even numbers; upper limit line segments are assigned odd num-
bers; see
CALCulate<Chn>:LIMit:LOWer[:DATA]
.
CALCulate<Chn>:LIMit:LOWer:STATe
does not affect
segments with odd numbers;
CALCulate<Chn>:LIMit:UPPer:STATe
does not
affect segments with even numbers.
R&S ZVR/ZVABT compatible commands