Command reference
R&S
®
ZNB/ZNBT
1487
User Manual 1173.9163.02 ─ 62
SOURce<Ch>:POWer<PhyPt>:CORRection:COLLect[:ACQuire]
<Sensor>
Initiates a source power calibration for the source port
<PhyPt>
using an external
power meter no. 1 or 2. To initiate a source power calibration for arbitrary power
meters, use the alternative commands listed in the program example below.
Note:
The command cannot be used unless a power meter is connected via GPIB bus,
USB or LAN interface and configured in the "External Power Meters" dialog.
Suffix:
<Ch>
.
Calibrated channel number
<PhyPt>
Calibrated port number
Setting parameters:
<Sensor>
ASENsor | BSENsor
ASENsor
–
external power meter Pmtr 1.
BSENsor
–
external power meter Pmtr 2.
*RST:
ASENsor
Example:
SOUR:POW3:CORR:COLL BSEN
Perform a source power calibration for port 3 using power meter
no. 2.
SOUR:POW:CORR:PMET:ID 2
Select power meter no. 2.
SOUR:POW:CORR:ACQ PORT,3
Perform a source power calibration for port 3 using the previ-
ously selected power meter no. 2.
Usage:
Setting only
Manual operation:
See
TRACe:CLEar
<MemTrace>
Deletes one of the memory traces
Mem<n>[Trc<m>]
, where n = 1, ... 8.
Setting parameters:
<MemTrace>
MDATa1 | MDATa2 | MDATa3 | MDATa4 | MDATa5 | MDATa6 |
MDATa7 | MDATa8
Identifier for the memory trace; see
Range:
MDATA<n> where <n> = 1 to 8.
Example:
SWE:POIN 20
Create a trace with 20 sweep points, making the created trace
the active trace of channel 1 (omitted optional mnemonic
SENSe1
).
TRAC:COPY "Mem_Pt20",CH1DATA
Copy the current state of the created trace to a memory trace
named "Mem_Pt20". The memory trace is not displayed.
DISP:WIND:TRAC2:FEED "MEM_PT20"
Display the created memory trace in the active diagram area
(diagram area no. 1).
R&S ZVR/ZVABT compatible commands