The notation for wave quantities follows the usual scheme of the vector network ana-
lyzer:
●
The a-waves are the outgoing/transmitted waves at the analyzer's test ports.
●
The b-waves are the incoming/measured waves.
●
The source port for the stimulus signal must be specified in addition.
●
The port number range covers all test ports of the analyzer.
Wave Quantity
Selects the type (left pull-down list) and the port number assignment (right pull-down
list) of the
.
Remote command:
CALCulate<Ch>:PARameter:MEASure
"<Trace_Name>", "A1" ...
CALCulate<Ch>:PARameter:SDEFine
"<Trace_Name>", "A1" ...
Use Generator as Source
If selected, the "Port" control lists the configured external generators. Otherwise it lists
the analyzer ports.
To configure external generators, use the "External Generators" dialog. See
ter 4.7.28, "External generators"
Source > Port
Selects the source for the stimulus signal. Depending on the state of the "Use Genera-
tor as Source" flag either analyzer ports ("Port 1" to "Port n") or external generators
("Gen 1", "Gen 2", ...) can be selected.
The analyzer places no restriction on the combination of source ports and port num-
bers of the measured wave quantity. For example, it is possible to measure a
2
while
port 1 is the source port (e.g. to estimate the directivity of the internal test set's direc-
tional element).
Remote command:
CALCulate<Ch>:PARameter:MEASure
"<Trace_Name>", "A1" ...
CALCulate<Ch>:PARameter:SDEFine
"<Trace_Name>", "A1" ...