GUI reference
R&S
®
ZNB/ZNBT
367
User Manual 1173.9163.02 ─ 62
patibility means that the "Sweep Type" of the two traces must match; the position and
number of the sweep points do not have to be the same.
The analyzer checks for compatibility before importing data. The "Select Parameter"
box remains empty if the selected file contains no compatible data.
"Import Complex Data" is a standard "Open File" dialog with an additional button.
Import Data to New Mem
Specifies whether the loaded data overwrite an existing memory trace, if available (box
unchecked), or whether they are used to generate a new memory trace (box checked).
If the box is unchecked and the active trace is a memory trace, then this memory trace
will be overwritten. If the box is unchecked and the active trace is a data trace, then the
data trace's last created memory trace will be overwritten (or a new memory trace will
be created, in case there was previously no memory trace assigned to this data trace).
Remote command:
5.5.11.2
Export Data - <File Type> dialog
The "Export Data - <File Type>" dialog stores data or memory traces to a trace file.
Trace files are ASCII files with selectable file format.
Access:
●
Trace – [Trace
Config] > "Trace Data" > "Export snp Files" – ...
●
Trace – [Trace
Config] > "Trace Data" > "Export" – ...
Data export can serve many purposes, e.g.:
●
To process and evaluate measurement data in an external application.
●
To store measurement data and reimport it in a future measurement session.
Background information
Refer to the following sections:
●
●
Chapter 4.4.2.3, "Finding the best file format"
Trace Config softtool