GUI reference
R&S
®
ZNB/ZNBT
397
User Manual 1173.9163.02 ─ 62
Background information
Refer to
"Basic marker search functions"
5.7.3.1
Controls on the Marker Search tab
Max / Min
Sets the active marker to the absolute maximum or minimum in the search range, i.e.
to the largest or smallest of all response values. If a complex trace format (e.g. a polar
diagram) is active, the marker is set to the measurement point with the maximum or
minimum magnitude.
"Max" and "Min" also overwrite the current "Search Mode" (--> "Search Min" and
"Search Max") and the "Peak Type" for the peak search functions.
Remote command:
CALCulate<Chn>:MARKer<Mk>:FUNCtion:EXECute MINimum | MAXimum
CALCulate<Chn>:MARKer<Mk>:REFerence:FUNCtion:EXECute MINimum |
MAXimum
CALCulate<Chn>:MARKer<Mk>:FUNCtion:RESult?
CALCulate<Chn>:MARKer<Mk>:REFerence:FUNCtion:RESult?
Center = Marker
See
"Center = Marker / Start = Marker / Stop = Marker / Span = Marker"
Next Peak
Sets the active marker to the next local maximum or minimum in the search range,
depending on the selected
.
Marker softtool