GUI reference
R&S
®
ZNB/ZNBT
505
User Manual 1173.9163.02 ─ 62
●
If inactive, the R&S
ZNB/ZNBT permanently refreshes all traces of all diagrams
which can put a heavy load on the connected switch matrices (if any).
Hence it is recommended (and default) to activate it, in particular if one of the matrices
uses
mechanical
switches (that wear off apart from making noise).
Apply
Is enabled when sufficient data have been acquired for the calibrated ports and stand-
ards and for the power meter. The button starts the calculation of the calibration data
and closes the calibration wizard. The current instrument settings are stored with the
correction data.
To avoid incompatibilities, older calibration data is deleted unless it has been transfer-
red into the "Cal Pool" using the "Calibration Manager".
Use the
to perform an additional source flatness calibration (see
Chapter 4.5.7.3, "Combining SMARTerCal with scalar power calibration"
Note: Checks during the calculation of correction data
Incompatibilities between the selected calibration type, the standards and the channel
settings can cause the calibration to be inaccurate. The analyzer auto-detects potential
sources of errors and displays appropriate, self-explanatory notice boxes.
Remote command:
[SENSe<Ch>:]CORRection:COLLect:SAVE:SELected[:DUMMy]
[SENSe<Ch>:]CORRection:COLLect:SAVE:SELected:DEFault
[SENSe<Ch>:]CORRection:COLLect:DELete
[SENSe<Ch>:]CORRection:DATA:PARameter<Sfk>?
[SENSe<Ch>:]CORRection:DATA:PARameter<Sfk>:PORT<PhyPt>?
[SENSe<Ch>:]CORRection:DATE?
[SENSe<Chn>:]CORRection:SSTate?
[SENSe<Ch>:]CORRection:STIMulus?
[SENSe<Ch>:]CORRection:STIMulus:PORT<PhyPt>?
5.11.1.11
SMARTerCal wizard (manual, MultiCal)
The multiple calibration version of the
guides you
through the setup and simultaneous manual acquisition of multiple SMARTerCals.
Except for the possibility to create multiple calibrations, the layout and functionality is
the same as for the single calibration version of the wizard.
Access:
Channel – [Cal] > "Start Cal" > "SMARTerCal"
–
"Start... (Manual)"
Background and related information
●
The "Multiple Cal in Calibration Wizard" feature is disabled by default and has to be
activated in the
of the "System Config" dialog.
●
Chapter 5.11.1.10, "SMARTerCal wizard (manual)"
on page 496 for more
information.
Cal softtool