GUI reference
R&S
®
ZNB/ZNBT
534
User Manual 1173.9163.02 ─ 62
All possible power calibrations can be performed from the
Port Overview
Shows all source ports together with their possible power calibrations. Unavailable cali-
brations (i.e. calibrations that have not been performed yet) are grayed out.
If external generators are configured, they appear as additional source ports G1 ... Gm
in the port overview. A source power calibration is available at each generator port.
Remote command:
SOURce<Ch>:POWer<PhyPt>:CORRection:STATe
SOURce<Ch>:POWer:CORRection:GENerator<Gen>[:STATe]
[SENSe<Ch>:]CORRection:POWer<PhyPt>[:STATe]
[SENSe<Ch>:]CORRection:PCAL
5.11.4.3
Calibration Manager dialog
The "Calibration Manager" dialog stores user correction data to the cal pool and
assigns stored correction data to channels.
See
Chapter 4.5.3, "Calibration pool"
on page 181 for background information.
Access:
Channel – [Cal] > "Use Cal" > "Cal Manager..."
Cal softtool