GUI reference
R&S
®
ZNB/ZNBT
541
User Manual 1173.9163.02 ─ 62
Note:
Calibration and port de-/activation.
The analyzer fimware automatically activates/deactivates ports during/after a (success-
ful) calibration:
●
Calibrated ports that were previously disabled, are automatically enabled as single-
ended logical ports.
●
An uncalibrated port that is not used by a measurement (i.e. the port is not required
by any trace of the related channel) is disabled.
Remote command:
The port parameters in many calibration commands define the calibrated ports.
Type
Selects the calibration type.
Remote command:
[SENSe<Ch>:]CORRection:COLLect:METHod:DEFine
<CalName>, MTOSm |
MUOSm | MOSM | MTRL, ...
Step 2: Connectors and Cal Kits
Connector/Gender
Specifies the connector and gender of the ports selected in
Note that METAS compatible file formats do not contain information about the related
connector type, so you have to make sure to choose a suitable
Remote command:
see
Cable
Allows you to specify which cables are connected to the related ports.
Cal softtool