GUI reference
R&S
®
ZNB/ZNBT
549
User Manual 1173.9163.02 ─ 62
New C Diagram
Creates a channel and a trace with default settings, and assigns the created trace to a
new diagram area.
The new channel is named Ch<n>, where <n> is the largest of all existing channel
numbers plus one. The name can be changed in the "Channel Manager" dialog.
Tip:
To create a channel and a trace with default settings and to display the trace in the
active
diagram area, use
Copy C Diagram
Copies the active channel with all its settings (including a possible channel calibration)
and traces, and displays the created traces in a new diagram area.
The new channel is named Ch<n>, where <n> is the largest of all existing channel
numbers plus one. The name can be changed in the "Channel Manager" dialog.
Tips:
Use
to display the new traces in an
existing
diagram area.
Delete Channel
Deletes the current channel including all the traces assigned to it and removes all dis-
play elements related to the channel from the diagram area. "Delete Channel" is dis-
abled if the recall set contains only one channel: In manual control, each recall set
must contain at least one diagram area with one channel and one trace.
Tips:
Use the "Undo" function to restore a channel that was unintentionally deleted.
Remote command:
OFF
Channel On
Toggles the measurement state of the
.
Remote command:
CONFigure:CHANnel<Ch>:MEASure[:STATe]
Fixture Simulator
The "Fixture Simulator" switch deactivates or activates the configured deembedding,
embedding, balanced ports, and port impedance settings for the selected channel.
When "Fixture Simulator" is
deactivated:
●
all balanced ports are resolved to single ended ports
●
all port impedances are set to default
●
all de/embeddings are disabled
At the GUI, the "Balanced Ports" dialog and the de/embedding tabs and dock widgets
are disabled. Related remote commands generate an error.
When "Fixture Simulator" is
reactivated
, the situation before the deactivation is
restored.
Note:
The "Offset" and "One Way Loss" settings are
not
affected by the "Fixture Simu-
lator" switch. Use the "All Compensation Off"/"All Compensation On" functions of the
"Fixture Compensation" dialog to de/activate these compensations as well (see
ter 5.13.2.2, "Fixture Compensation dialog"
Remote command:
CALCulate<Ch>:TRANsform:VNETworks:FSIMulator[:STATe]
Channel Config softtool