GUI reference
R&S
®
ZNB/ZNBT
623
User Manual 1173.9163.02 ─ 62
If the measured quantity is a ratio, or if it is derived from a ratio, its receiving port is
given as the index of the wave quantity in the numerator. If the active trace shows an
S-parameter S
ij
, then "Auto Length" adds a length offset at port i. If the receive port is
balanced, then the same offset and length is applied to both its physical ports.
Remote command:
[SENSe<Ch>:]CORRection:LOSS:AUTO
Fixture Compensation...
The "Fixture Compensation..." button opens the
.
Evaluation Range...
The "Evaluation Range..." button opens the
that allows you to
restrict the frequency range of auto length (and loss) calculations.
5.13.4
Single Ended tab
Allows you to specify 2-port deembedding/embedding networks for each physical port.
Such a network is either defined:
●
via its S-parameters stored in a two-port Touchstone file (
*.s2p
) or
●
by selecting a predefined lumped element model and specifying the available
parameters (resistances/conductances, capacitances, inductances)
See
Chapter 4.6.2.3, "Circuit models for 2-port networks"
Use the complementary dock widget to activate or deactivate dembedding/embedding
for selected ports (see
Chapter 5.13.1.3, "Single Ended panel"
Background information
Chapter 4.6.2.10, "Combining Several de-/embedding networks"
Offset Embed softtool