GUI reference
R&S
®
ZNB/ZNBT
758
User Manual 1173.9163.02 ─ 62
Remote command:
Show Info Messages / Show Warning Messages / Show Error Messages
Selectively disables/enables display of information popups for the related event type.
If information popups are globally switched off (
unchecked), these buttons are disabled. However, their checked state is memorized.
Remote command:
SYSTem:ERRor:DISPlay:INFO
SYSTem:ERRor:DISPlay:WARNings
SYSTem:ERRor:DISPlay:ERRor
Show Remote Error Info Messages
Defines whether information popups are displayed whenever a remote control com-
mand error occurs.
The displayed information can be useful for program development and optimization; it
does not necessarily indicate that a remote control script is faulty or non-executable.
Note
●
If either "Show Instrument Messages" or "Show Info Messages" is unchecked, this
button is disabled. However, its checked state is memorized.
●
For SCPI error
-113, Undefined header
, no tooltip is displayed.
Remote command:
Channel Bits tab
Sets a channel-dependent 8-bit decimal value (0, ..., 255) to control eight independent
output signals at the User Port connector (lines 8 to 11 and 16 to 19).
Setting the channel bits does not change the analyzer state.
Channel Bits (Decimal)
Entry of the 8-bit decimal value (0 ... 255) for the selected channel. The channel bits
control eight output signals at the User Port connector. The signals are 3.3 V TTL sig-
nals which can be used to differentiate between up to 256 independent analyzer
states. For an application example, refer to the description of the remote-control com-
mand.
Setup softtool