Configured Devices
Table with all generators in use with their properties. The properties of manually config-
ured generators ("Add Device", opens the "Add External Generator" dialog) may be
changed in the dialog.
The following generator properties are not defined in the "Add External Generator" dia-
log but configurable here:
●
"Ext. Ref." switches the generator to either external (box checked) or internal fre-
quency reference. The setting does not affect the frequency reference settings of
the analyzer System – [Setup] > "Freq. Ref.". Make sure to establish consistent
settings in your test setup (one instrument is the master, the others use external
reference frequency). See
●
"Fast Sweep" enables or disables the fast sweep mode for external generators that
support a frequency and level list mode (triggered mode). See
●
"Sweep End" defines the output power of the external generator after the end of a
sweep or sweep sequence (single sweep mode; Channel – [Sweep] > "Sweep
Control" > "Single"). The minimum power value ("Switch to Min. Power") depends
on the generator type. "Set to Selected Power" means that the generator is com-
manded to transmit at the selected "End Power".
The following symbols (grayed out for
used
devices) indicate the status of the respec-
tive device:
●
–
The device is online (connected, switched on, ready to be used).
●
–
There is a problem with the device.
This state can be caused by different problems:
–
General communication error
In this case, check whether the device is properly connected to the configured
interface.
–
Self test error
In this case enable error logging for external devices (see
for self test error codes of the
device.
●
–
The device driver does not match exactly the device type. Probably the device
cannot be controlled properly. Measurements results are questionable.
Remote command:
SYSTem:COMMunicate:RDEVice:GENerator<Gen>:DEFine
SYSTem:COMMunicate:RDEVice:GENerator:DELete
SYSTem:COMMunicate:RDEVice:GENerator<Gen>:CATalog?
SYSTem:COMMunicate:RDEVice:GENerator<Gen>:DEFine
SYSTem:COMMunicate:RDEVice:GENerator<Gen>:SEPMode
SYSTem:COMMunicate:RDEVice:GENerator<Gen>:SEPower