Command reference
R&S
®
ZNB/ZNBT
896
User Manual 1173.9163.02 ─ 62
CALCulate:FMODel:DELT:FREQuencies:USEDefault
<Boolean>
Delta-L computes the uncertainty in the loss-per-inch calculation at specified frequen-
cies.
This command specifies whether to use the tool's default frequencies or custom fre-
quencies, which can be defined using
CALCulate:FMODel:DELT<Ph_pt>:FRE
commands.
Parameters:
<Boolean>
ON (1)
Use default frequencies
OFF (0)
Use custom frequencies
Options:
R&S
ZNB/ZNBT-K231
Manual operation:
See
CALCulate:FMODel:DELT:FREQuencies:USER?
The Delta-L algorithm can calculate the loss-per-inch uncertainties either at default or
at custom frequencies (see
CALCulate:FMODel:DELT:FREQuencies:
).
This command returns the custom frequencies, defined using
etc.
Usage:
Query only
Options:
R&S
ZNB/ZNBT-K231
Manual operation:
See
CALCulate:FMODel:DELT:FREQuencies:USER:ADD
<UserFrequencies>
CALCulate:FMODel:DELT:FREQuencies:USER:DELete
<Frequency>
The Delta-L algorithm can calculate the loss-per-inch uncertainties either at default or
at custom frequencies (see
CALCulate:FMODel:DELT:FREQuencies:
).
CALCulate:FMODel:DELT<Ph_pt>:FREQuencies:USER:ADD|DELete
adds/
deletes a frequency to|from the list of custom frequencies.
Setting parameters:
<Frequency>
Custom frequency
Range:
The frequency range of the analyzer
Default unit: Hz
Usage:
Setting only
Options:
R&S
ZNB/ZNBT-K231
Manual operation:
See
SCPI command reference