Command reference
R&S
®
ZNB/ZNBT
926
User Manual 1173.9163.02 ─ 62
Suffix:
<Chn>
.
Channel number used to identify the active trace
Parameters:
<StartStim>
<StopStim>
Pairs of stimulus values, each pair confining a limit line segment.
See also
Chapter 4.4.1.1, "Rules for limit line definition"
on page 145.
If not specified, the units are adjusted to the sweep type of the
active channel (
).
Range:
Virtually no restriction for limit segments.
*RST:
A segment that is created implicitly, e.g. by means
of CALCulate<Ch>:LIMit:UPPer[:DATA] or CALCu-
late<Ch>:LIMit:LOWer[:DATA,], covers the maxi-
mum sweep range of the analyzer.
Example:
*RST; :CALC:LIM:CONT 1 GHZ, 2 GHZ
Select a lin. frequency sweep (default) and define an upper limit
line segment in the stimulus range between 1 GHz and 2 GHz,
using default response values (
–
40 dB).
CALC:LIM:DISP ON
Show the limit line segment in the active diagram.
Manual operation:
See
CALCulate<Chn>:LIMit:CONTrol:SHIFt
<LimShift>
Shifts an existing limit line in horizontal direction. See also
Suffix:
<Chn>
.
Channel number used to identify the active trace
Setting parameters:
<LimShift>
Offset value for the limit line
Range:
Virtually no restriction for limit segments
Default unit: NN
Example:
*RST; :CALC:LIM:CONT 1 GHZ, 2 GHZ
Define a limit line segment in the stimulus range between 1 GHz
and 2 GHz, using default response values.
CALC:LIM:CONT:SHIF 1; :CALC:LIM:CONT?
Shift the segment by 1 Hz. The modified limit line segment
ranges from
1000000001
(Hz) to
2000000001
(Hz).
Usage:
Setting only
Manual operation:
See
SCPI command reference