Command reference
R&S
®
ZNB/ZNBT
938
User Manual 1173.9163.02 ─ 62
Suffix:
<Chn>
.
Channel number used to identify the active trace
<Seg>
Segment number
Parameters:
<LimLineType>
LMIN | LMAX | OFF
Limit line type
Range:
LMAX (upper limit line segment), LMIN (lower limit
line segment), OFF (limit check switched off, limit
line segment not deleted)
*RST:
LMAX
Example:
*RST; :CALC:LIM:UPP 0, 0
Define an upper limit line segment across the entire sweep
range, using a constant upper limit of 0 dBm.
CALC:LIM:SEGM:TYPE LMIN
Turn the defined limit line segment into a lower limit line seg-
ment.
Manual operation:
See
CALCulate<Chn>:LIMit:SOUNd[:STATe]
<Boolean>
Deprecated command
: Since version 3.45, generating sounds is no longer supported
by the analyzer firmware.
Switches the acoustic signal (fail beep) on or off. The fail beep is generated each time
the analyzer detects an exceeded limit.
Suffix:
<Chn>
.
Channel number used to identify the active trace
Parameters:
<Boolean>
ON | OFF - Fail beep on or off.
*RST:
OFF
Example:
CALC:LIM:STAT ON; SOUN ON
Switch the limit check on and activate the fail beep.
CALCulate<Chn>:LIMit:STATe
<Boolean>
Switches the limit check (including upper and lower limits) on or off.
Tip:
Use
CALCulate<Ch>:LIMit:UPPer:STATe
or
CALCulate<Ch>:LIMit:LOWer:STATe
to switch on or off the individual limit checks
for upper or lower limit lines.
Suffix:
<Chn>
.
Channel number used to identify the active trace
SCPI command reference