Command reference
R&S
®
ZNB/ZNBT
950
User Manual 1173.9163.02 ─ 62
CALCulate<Chn>:MARKer<Mk>:FUNCtion:DOMain:USER:STARt
<StarSearchRange>
CALCulate<Chn>:MARKer<Mk>:FUNCtion:DOMain:USER:STOP
<StopSearchRange>
CALCulate<Chn>:MARKer<Mk>:REFerence:FUNCtion:DOMain:USER:STARt
<StarSearchRange>
CALCulate<Chn>:MARKer<Mk>:REFerence:FUNCtion:DOMain:USER:STOP
<Value>
These commands define the start and stop values of the search range selected via
CALCulate<Chn>:MARKer<Mk>:FUNCtion:DOMain:USER[:RANGe]
or
CALCulate<Chn>:MARKer<Mk>:REFerence:FUNCtion:DOMain:USER[:RANGe]
.
Suffix:
<Chn>
.
Channel number used to identify the active trace
<Mk>
Marker number (ignored for reference markers)
Parameters:
<Value>
Beginning or end of the search range.
Range:
Maximum allowed sweep range, depending on the
instrument model and on the sweep type.
Default unit: NN
Example:
See
CALCulate<Chn>:MARKer<Mk>:FUNCtion:DOMain:
Manual operation:
See
CALCulate<Chn>:MARKer<Mk>:FUNCtion:EXECute
[<SearchMode>]
CALCulate<Chn>:MARKer<Mk>:REFerence:FUNCtion:EXECute
[<SearchMode>]
Selects a search mode for the related marker (see
and executes the search.
Except for bandfilter searches (
BFILter
), you have to create the related marker using
CALCulate<Chn>:MARKer<Mk>[:STATe] ON
or
MARKer<Mk>:REFerence[:STATe] ON
before using this command.
Suffix:
<Chn>
.
Channel number used to identify the active trace
<Mk>
Marker number
Ignored for reference markers and for bandfilter searches
(
BFILter
), which always use markers M1 to M4.
Setting parameters:
<SearchMode>
MAXimum | MINimum | RPEak | LPEak | NPEak | TARGet |
LTARget | RTARget | BFILter | MMAXimum | MMINimum |
SPRogress
See list of search modes below.
SCPI command reference